Sunday, January 24, 2010

Nipro Medical Corporation Issues a Voluntary Recall of All GlucoPro Insulin Syringes

Nipro Medical Corporation, Miami FL, is initiating a nationwide recall of all GlucoPro Insulin Syringes (This does not include the GlucoPro syringe specific for use with the Amigo Insulin pump). These syringes may have needles that detach from the syringe. If the needle becomes detached from the syringe during use, it can become stuck in the insulin vial, push back into the syringe, or remain in the skin after injection.

Consumers who have GlucoPro Insulin Syringes should stop using and return them to point of sale for reimbursement.

This recall includes all product codes and lot numbers with expiration dates before 2011-11 (Nov 1, 2011).

Product Code Lot # Expiry date
JD+01U3008-5C A08022 2011-11
JD+01U3013-5C A08013 2011-06
JD+01U3013-5C A08017 2011-08
JD+01U3108-5C A08013 2011-06
JD+01U3108-5C A08017 2011-08
JD+03U3008-5C C08022 2011-11
JD+03U3013-5C C08013 2011-06
JD+03U3013-5C C08017 2011-08
JD+03U3108-5C C08013 2011-06
JD+03U3108-5C C08017 2011-08
JD+05U3008-5C B08022 2011-11
JD+05U3013-5C B08013 2011-06
JD+05U3013-5C B08017 2011-08
JD+05U3108-5C B08013 2011-06
JD+05U3108-5C B08017 2011-08


The firm voluntarily recalled the products after learning of the possibility of needle detachment. FDA has been apprised of this action.

No injuries have been reported to date.

Product was distributed nationwide, including Puerto Rico.

Company is notifying its distributors and customers by Fax and Email and is arranging for return of all recalled products.

Consumers with questions may contact the company at 305.599.7174 x249.

Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

* Online: www.fda.gov/medwatch/report.htm
* Regular Mail: use postage-paid FDA form 3500 available at: www.fda.gov/MedWatch/getforms.htm.
Mail to MedWatch 5600 Fishers Lane, Rockville, MD 20852-9787
* Fax: 1-800-FDA-0178

-----
www.fayettefrontpage.com
Fayette Front Page
www.georgiafrontpage.com
Georgia Front Page
Follow us on Twitter: @GAFrontPage

Wednesday, October 28, 2009

Qualitest Pharmaceuticals Issues a Voluntary Nationwide Recall of All Accusure® Insulin Syringes

Huntsville AL-Qualitest Pharmaceuticals today issued a voluntary nationwide recall of all Accusure® Insulin Syringes. The distributed syringes are of the following descriptions and NDC numbers: 28G 1/2cc, NDC 0603-6995-21;28G 1cc, NDC 0603-6996-21; 29G 1/2cc NDC 0603-6997-21, 29G 1cc, NDC 0603-6998-21, 30G 1/2cc, NDC 0603-999-21, 30G 1cc, NDC 0603-7000-21, 31G 1/2cc, NDC 0603-7001-21; and 31G 1cc, NDC 0603-7002-21. All Accusure® Insulin Syringes regardless of lot number are subject to this recall. These syringes were distributed between January 2002 and October 2009 to wholesale and retail pharmacies nationwide (including Puerto Rico). The syringes in these lots may have needles which detach from the syringe.

If the needle becomes detached from the syringe during use, it can become stuck in the insulin vial, push back into to the syringe, or remain in the skin after injection.

Consumers who have any Accusure® Insulin Syringes should stop using them and contact Qualitest at 1-800-444-4011 for reimbursement. You can find the lot number on the white paper backing of each individual syringe.

Qualitest is notifying all customers who received these syringes and arranging for the return of any affected product.

The recall is being made with the knowledge of the Food and Drug Administration.

Consumers with questions may contact Qualitest at 1-800-444-4011 for more information.

Adverse reactions or quality problems experienced with the use of this product may be reported to FDA's MedWatch Adverse Event Reporting program either on line, by regular mail or by fax.

*
Online:www.fda.gov/medwatch/report.htm
Regular Mail: Use postage-paid FDA form 3500 available at: www.fda.gov/medwatch/getforms.htm
Mail to MedWatch 5600 Fishers Lane, Rockville, MD 20852-9787
Fax: 1-800-FDA-0178

-----
www.fayettefrontpage.com
Fayette Front Page
www.georgiafrontpage.com
Georgia Front Page

Wednesday, November 5, 2008

FDA Reports Nationwide Recall of Mislabeled ReliOn Insulin Syringes

The U.S. Food and Drug Administration is notifying health care professionals and patients that Tyco Healthcare Group LP (Covidien) is recalling one lot of ReliOn sterile, single-use, disposable, hypodermic syringes with permanently affixed hypodermic needles due to possible mislabeling. The use of these syringes may lead to patients receiving an overdose of as much as 2.5 times the intended dose, which may lead to hypoglycemia, serious health consequences, and even death.

The recall applies to the following lot number and product information:

-- Lot Number 813900
-- ReliOn 1cc, 31-gauge, 100 units for use with U-100 insulin

Only ReliOn syringes from this lot number and labeled as 100 units for use with U-100 insulin are the subject of the recall.

These syringes are distributed by Can-Am Care Corp and sold only by Wal-Mart at Wal-Mart stores and Sam's Clubs under the ReliOn name. Wal-Mart requests that all users of ReliOn 31-gauge, 1cc syringes return those labeled as 100 units for use with U-100 insulin from Lot Number 813900 to their local Wal-Mart store or Sam's Club pharmacy. Customers will be provided with replacement product.

The FDA urges patients and health care professionals to check their syringe packaging carefully for syringes labeled as 100 units for use with U-100 insulin from Lot Number 813900.

Consumers and health care professionals who suspect they have the recalled product may also contact Covidien at 866-780-5436 or www.relion.com/recall for more information.

ReliOn Insulin Syringes consist of a syringe barrel, a plunger rod, and a hypodermic needle attached to the tip of the syringe.

During the packaging process for this lot, some syringes labeled for use with U-40 insulin were mixed with syringes labeled for use with U-100 insulin, then all packaged individually and in boxes as 100 units for use with U-100 insulin.

The manufacturer has distributed 4,710 boxes in the recalled lot, which equals 471,000 individual syringes. Wal-Mart sold the syringes at Wal-Mart stores and Sam's Clubs from Aug. 1, 2008, until Oct. 8, 2008.

Tyco Healthcare Group LP (Covidien) voluntarily recalled this lot of syringes on Oct. 9, 2008, asking that any units of the affected product be removed from inventory and placed in quarantine. Wal-Mart posted the recall announcement in Wal-Mart stores and Sam's Clubs, as well as on its Web site, and sent letters to more than 16,500 customers notifying them of the recall.

The manufacturer has received one adverse report related to a syringe from this product lot.
Health care professionals and consumers may report serious adverse events (side effects) or product quality problems with the use of this product to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail, fax or phone.

--Online: www.fda.gov/MedWatch/report.htm
--Regular Mail: use postage-paid FDA form 3500 available at: www.fda.gov/MedWatch/getforms.htm and mail to MedWatch, 5600 Fishers Lane, Rockville, MD 20852-9787
--Fax: (800) FDA-0178
--Phone: (800) FDA-1088

-----